Degassing pump drive, Fresenius 4008B/H/S and 4008S V10 (6707671) — Image

Degassing pump drive, Fresenius 4008B/H/S and 4008S V10 (6707671)

Complete degassing pump drive for the Fresenius 4008B, 4008H, 4008S and 4008S V10.

KES49,700.00

Description

Part number6707671
FitsFresenius 4008B, 4008H, 4008S and 4008S V10
ConditionNew, genuine OEM
AvailabilityStocked in Nairobi
UnitEach

Before you order

Match the part number on the old part or in the machine's spare parts list. Not sure? Send a photo of the old part and its label on WhatsApp or to med@nanto.com and we confirm it before dispatch.

Service notes

Degassing pressure that will not hold, or a drive that is noisy or stalls, are common signs. The 4008 technical manual gives a rated degassing pressure of minus 0.81 to minus 0.85 bar at port D.

Check the filter first: the manual lists a clogged filter upstream of the degassing pump (filter 210, or the cartridge filter) as a cause of the same errors. Your biomedical engineer should confirm the fault before you order.

Read more: how to tell the degassing drive is failing.