Degassing pump drive, Fresenius 4008B/H/S and 4008S V10 (6707671)
Complete degassing pump drive for the Fresenius 4008B, 4008H, 4008S and 4008S V10.
Description
| Part number | 6707671 |
|---|---|
| Fits | Fresenius 4008B, 4008H, 4008S and 4008S V10 |
| Condition | New, genuine OEM |
| Availability | Stocked in Nairobi |
| Unit | Each |
Before you order
Match the part number on the old part or in the machine's spare parts list. Not sure? Send a photo of the old part and its label on WhatsApp or to med@nanto.com and we confirm it before dispatch.
Service notes
Degassing pressure that will not hold, or a drive that is noisy or stalls, are common signs. The 4008 technical manual gives a rated degassing pressure of minus 0.81 to minus 0.85 bar at port D.
Check the filter first: the manual lists a clogged filter upstream of the degassing pump (filter 210, or the cartridge filter) as a cause of the same errors. Your biomedical engineer should confirm the fault before you order.
Read more: how to tell the degassing drive is failing.